Defibrillator Model
|
With ECG M3860A. With Text Only Display M3861A.
|
|
How Supplied
|
Defibrillator, instructions for use, standard battery, defibrillator pads (2 pair), quick reference card.
|
|
Waveform
|
Truncated exponential biphasic. Waveform parameters adjusted as a function of patient impedance.
|
|
Energy
|
Single energy output. Adults:150 J nominal into a 50 ohm load. Pediatric: 50 J nominal into a 50 ohm load.
|
|
Charge Time From Shock Advised
|
Typically less than 10 seconds.
|
|
Shock-to-Shock Cycle Time
|
Typically less than 20 seconds (including analysis time) in automated mode.
|
|
Protocol
|
Text and voice prompts guide user through protocol. Follows pre-configured settings. Can be modified with the M3864A training & administration pack.
|
|
Shock Delivery
|
Via defibrillator pads placed in anterior-anterior (lead II) position for adult defibrillation and anterior-posterior for infant/child defibrillation.
|
|
Controls
|
On/off, shock, screen contrast/option buttons.
|
|
Indicators
|
LCD screen, beeper, audio speakers, status indicator, shock button, connector socket LED.
|
|
Advanced Mode
|
Configurable protocol.
|
Size Height Width Depth
|
6.6 cm. (2.6 inches). 21.8 cm. (8.6 inches). 21.8 cm. (8.6 inches).
|
Weight With battery Without battery
|
2.1 kg. (4.7 pounds). 1.8 kg. (3.9 pounds).
|
|
Screen
|
High-resolution LCD with bright back light.
|
|
Screen Dimensions
|
2.8 inches wide x 2.3 inches high (7.0 cm x 5.8 cm).
|
|
Display Range
|
Differential: ±2 mV full scale (nominal).
|
|
Sweep Speed
|
23 mm/second (nominal).
|
|
Frequency Response
|
1 Hz to 20 Hz (-3dB) (nominal).
|
|
Sensitivity
|
1.16 cm/mV (nominal).
|
|
Heart Rate
|
30 to 300 beats per minute updated each analysis period during monitoring.
|
|
Monitored Lead
|
Anterior-anterior (lead II) placement with adult defibrillation pads or ECG Assessment module (M3860A only).
|
|
Patient Analysis
|
Per protocol, evaluates patient ECG and signal quality to determine if a shock is appropriate, and evaluates connection impedance for proper defibrillation pad contact.
|
|
Sensitivity/Specificity
|
Meets AAMI DF-39 guidelines.
|
|
Type
|
12 VDC 4.2 Ah lithium manganese. Disposable, recyclable, long-life, primary cells.
|
|
Capacity
|
Minimum 300 shocks or 12 hours operating time.
|
|
Install-By Date
|
Battery is labeled with an install by date at least 5 years from date of manufacture.
|
|
Standby Life
|
Defines how long the battery will power the AED in standby operation within the standby temperature range (one battery insert test and no uses). 4 years minimum when battery is installed by the Install-By-Date (5 years typical).
|
|
Configuration
|
DP2: two pair or DP6: six pair.
|
|
How Supplied
|
Disposable self-adhesive defibrillation pads with integrated cable and connector.
|
|
Surface Area
|
100 cm2 each (nominal active surface area).
|
|
Cable Length
|
Approximately 122 cm (48 inches).
|
|
Patient
|
Under 8 years or 25 kg (55 pounds).
|
|
Defibrillator Compatibility
|
FR2-series (FR2 and FR2+) automated external defibrillator only.
|
|
Configuration
|
M3870A FR2 Infant/Child Reduced Energy Defibrillator Pads. 1 set per package.
|
|
How Supplied
|
Disposable self-adhesive pads with cable and connector.
|
|
Energy Delivered
|
Reduces defibrillator shock energy to nominal 50 Joules into a 50 ohm load.
|
|
Surface Area
|
44 cm2 each pad (nominal active surface area).
|
|
Cable Length
|
Approximately 122 cm (48 inches).
|
|
Standard Event Review
|
Elapsed time and number of shocks are displayed on screen.
|
|
Enhanced Event Review
|
Optional data card (M3854A) expands the above on-screen event review capabilities. Review chronological events in detail including ECG. 8 Hours of event & ECG recording or one hour if voice recording is activated.
|
|
Sealing
|
Meets IEC529 class IP54 with battery and data card tray installed.
|
|
Temperature
|
Operating: 0º - 50º C (32º - 122º F). Standby: 0º - 43º C (50º - 109º F). Standby applies to AED with battery installed and stored with defibrillation pads.
|
|
Humidity
|
Operating: 0% to 95% relative humidity (non-condensing). Standby: 0% to 75% relative humidity (non-condensing).
|
|
Altitude
|
-500 to 15,000 feet per MIL-STD-810E 500.3 Procedure II.
|
|
Aircraft
|
Method: RTCA/DO-160D: 1997 Section 21 (Category M -Charging).
|
|
Shock/Drop Abuse Tolerance
|
1 meter any edge, corner or surface. MIL-STD-810E 516.4 Procedure IV.
|
|
Vibration
|
MIL-STD-810E 514.4-17.
|
|
EMI
|
Requirements: Tested to EN60601-1-2 Limits. Radiated: Method EN55011: 1998 Group 1 Level B. Immunity: Method EN61000-4-3:1998 Level 2.
|
|
Automatic Self-Tests
|
Tests internal circuitry, waveform delivery system, and battery capacity. Verifies calibration of key circuits monthly.
|
|
Automatic Self-Test Frequency
|
Daily when stored within operating environmental conditions.
|
|
Status Indication
|
Dynamic visual and audible indication of self-test results. Indicates device readiness.
|
|
Battery Insertion Test
|
Upon battery insertion, extensive automatic self-tests and user inter-active tests check device readiness. Verifies calibration of key circuits.
|
|
Automatic Standby Temperature Monitoring
|
Instrument automatically monitors temperature and warns user if device is stored outside of standby temperature range.
|
|
Function
|
Places FR2+ in scenario-based training mode and disables energy delivery. 10 real-world scripts provided. Permits modification of preprogrammed FR2+ protocol.
|
|
Type
|
12 VDC 1.1 Ah Rechargeable Nickel Metal Hydride.
|
|
Capacity
|
Minimum 4 hours training time.
|
|
Recharge Time
|
90 minutes to full capacity using M3855A charger (sold separately).
|