Welch Allyn - AED 10 Technical Specifications

DEFIBRILLATOR
Output: Welch Allyn Orbital™ Biphasic Truncated exponential.
Energy Select: 200, 300, 360J or user configurable among the energy options: 150, 200, 300, 360
Charge Time: Less than 5 seconds to 150J, less than 15 seconds to 360J.
Analysis Time: 4-16 sec.
Audible Prompts: 18 audible prompts.
Visual Prompts: 20 text screen prompts.
Controls: Four buttons - On/Off, Shock, and 2 software configurable buttons.
Accuracy: ±10% or 1J (whichever is greater) at 50Ω.

DISPLAY
Size: 54 x 36 mm.
Type: Backlit monochrome LCD
Display Resolution: 160 x 100 pixels.
Low Battery Indicator: Battery Icon gauge on display with 10 capacity levels.
Backlight: LED.

BATTERY
Non-Rechargeable: LiMnO2 12V, 2.6 Ah.
Capacity: 90 discharges at 360 Joules or 125 discharges at 200 Joules or 150
discharges at 150 Joules or 5 hours ECG monitoring.
Shelf Life: (25°C±15°C) 10 years (5 years storage + 5 years standby); 5 years standby
(after installation).*
* Capacity may be diminished at operating temperature extremes.

GENERAL
Weight: 3.5 lbs. with battery (1.6 kg)
Size: 8.25" x 6.875" x 2.75" (210 x 175 x 70 mm)
Operating Temperature: 0° to 50°C
Storage Temperature: -30° to 70°C (without battery)
Humidity: Up to 95% or 10° to 95°C (non-condensing)
Altitude: Up to 15,000 feet (-150 to 4570 m)
Shock: Mil Std 810E method 516.4, procedure 1 (40G, 6-9ms pulse, 1/2 sine each axis)
Vibration: Mil Std 810 method 514.4, category 1
Water resistance: IEC 529 IPX4
Drop Test: Mil Std 810F method 516.5 Procedure IV (1 meter drop onto any edge, corner, or surface)

ELECTRICAL ISOLATION
Power: Unit operates on internal battery only.
External Electrical Connections: No external devices are attached to the unit.
Risk Current Category: Class II equipment with defibrillator-proof BF type patient applied parts (as per definition of IEC 60601-1 standard.)

ELECTROMAGNETIC COMPATIBILTY
Radiated Emissions: EN55011; CISPR11 B
ESD: EN61000-4-2; 8KV air 6KV contact
Radiated Susceptibility: EN61000-4-3 (10 V/M (20 v/m EN60601-2-4))
Magnetic Field Emissions: MIL STD RE101 (AAMI DF 24.3.18.1)
Magnetic Field Susceptibility: MIL STD RS101 1 Gauss, 47 Hz-1.8 KHz.

DATA MANAGEMENT
Event Documentation: Internal.
Internal Memory Capacity: 1MB: 250 4-sec ECG samples or 3000 time stamped events.
Quick Report: Treatment Summary, Event Log, Test Log.
Communication: RS-232 via Infrared (IR) port to PC or printer.

There are no products in your cart.

Lean More about AED Headquarters

Years in Business

©2004-2010 Save A Life, Incorporated. All rights reserved.
Access to and use of this Web Site is subject to these
Terms of Use.
™TradeMark Registrations are currently pending with Federal Government

links:popar hair wigs Hair pieces wigs uk watch toy sale toywatch uk toy watch uk moncler moncler jacket moncler coat Moncler ugg uk ugg boots sale ugg boots hair wigs Hair pieces wigs uk watch toy sale toywatch uk toy watch uk moncler moncler jacket moncler coat Moncler ugg uk ugg boots sale ugg boots